FLUKE 787 : *****TRUE RMS MULTIMETER* : Fluke : Nedis
Click here to go to the Nedis website.


- 60.000 products with more info
- more data,more search engines
- price and stock information
- sort articles on popularity
- quick ordering
- special offers
- new in the program
- order history
- open invoices
- backorder overview


FLUKE 787


*****TRUE RMS MULTIMETER*








The Fluke 787 ProcessMeter™ combines a DMM and a Loop Calibrator in one rugged, handheld tool - for about what you would expect to pay for a loop calibrator alone. The 787 starts with the Fluke 87 DMM you know and trust, then adds ability to measure, source, and simulate DC loop current with 0.05% accuracy and 1 micro-amp resolution. When you carry the ProcessMeter you`re ready for any loop troubleshooting or calibration task. And since it`s dust- and splash-resistant and EMI-shielded, the 787 can survive in your tool belt.


· DMM and Loop Calibrator in one tool
· Precision 1000V, 440 mA Digital Multimeter
- Measure AC and DC volts, AC and DC current, resistance, continuity, and frequency
- 0.1% DC voltage accuracy
- 0.05% DC current accuracy, 1 uA resolution to 30 mA
- Simultaneous mA and % of scale readout
- True RMS AC voltage measurement to 1 kHz
- Frequency measurement to 20 kHz
- Min/Max/Average/Hold/Relative modes
- Diode Test and Continuity Beeper
· DC current source/loop calibrator
- 20 mA current source/simulator
- Simultaneous mA and % of scale readout
- Manual Step (25%, Coarse, Fine) plus Auto Step and Auto Ramp
- Clear LCD with backlight; 4,000 counts (30,000 counts for DC current)
· Externally accessible battery for easy battery changes
· 1000V overload protection on V, ohms, frequency
· 150V overload protection on mA, backed-up by 440 mA 1000V fuse
· Designed to meet 1000 volt EN 61010 Cat III standards


Function Range Resolution Accuracy ±( % of reading+LSD)

DC Voltage Measurement
4.000 V DC 0.001 V 0.1% + 1
40.00 V DC 0.01 V 0.1% + 1
400.0 V DC 0.1 V 0.1% + 1
1000 V DC 1 V 0.1% + 1

DC milliVolts Measurement
400.0 mV DC 0.1 mV 0.1% + 1

AC Voltage Measurement "(true-rms responding," specified from 5% to 100% of range)
50 Hz-60 Hz 45 Hz-500 Hz 500 Hz-1 kHz
400 mV 0.1 mV ±(0.7%+4) ±(1.2%+4) ±(5.0%+4)
4.000 V 0.001 V ±(0.7%+2) ±(1.2%+4) ±(5.0%+4)
40.00 V 0.01 V ±(0.7%+2) ±(1.2%+4) ±(5.0%+4)
400.0 V 0.1 V ±(0.7%+2) ±(1.2%+4) ±(5.0%+4)
1000 V 1 V ±(0.7%+2) ±(1.2%+4) ±(5.0%+4)

AC Current (45 Hz to 2 kHz)
0.440 A 0.001 A ±(1% + 2)

DC Current 30.000 mA
0.001 mA ±(0.05% + 2)
0.440 A 0.001 A ±(0.2% + 2)

Ohms Measure
400.0 Ohms 0.1 Ohms 0.2% + 2
4.000 kOhms 0.001 kOhms 0.2% + 1
40.00 kOhms 0.01 kOhms 0.2% + 1
400.0 kOhms 0.1 kOhms 0.2% + 1
4.0 MOhms 0.001 MOhms 0.35% + 3
40 MOhms 0.01 MOhms 2.5% + 3

Frequency Measure (0.5 Hz to 20 kHz)
199.99 Hz 0.01 Hz ±(0.005% + 1)
1999.9 Hz 0.1 Hz ±(0.005% + 1)
19.999 kHz 0.001 kHz ±(0.005% + 1)